EasyManua.ls Logo

Sirona Intego - Sterilize;Disinfect the Suction Handpieces

Sirona Intego
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 498 D3543
D3543.201.01.05.02 08.2016
245
Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.4Vacuum system
båÖäáëÜ
5.4.4 Sterilize/disinfect the suction handpieces
Sterilization/disinfection
All parts of the suction handpieces can be sterilized and thermally
disinfected.
Lubricating suction handpieces
The disconnection points of the suction handpieces must be lubricated
once a week and after each thermal disinfection or sterilization process.
Grease
No surgical suction device option
1. Detach the handpiece of the spray aspirator and the saliva ejector
from the suction hoses.
2. Detach the rotary joint of the handpiece from the spray aspirator.
3. Regrease the disconnection points and O rings of the handpieces.
E
D
C
B
A
134°C 274°F
3 min
F
ASuction tip
B Suction Handpiece
C Rotary joint
D Saliva ejector handpiece
E Surgical suction tip
F Intermediate piece
CAUTION
Unsuitable greases
Greases that are not food safe may endanger the patient's health
Rubber materials such as e.g. O-rings are corroded by unsuitable
greases.
Do not under any circumstances use vaseline or silicone greases.
Use only greases approved by Sirona.

Table of Contents

Other manuals for Sirona Intego

Related product manuals