64 57 498 D3543
D3543.201.01.05.02 08.2016
249
Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.5Components of the water unit
båÖäáëÜ
5.5.2 Clean/disinfect the cuspidor
The cuspidor bowl, the tumbler holder A and the tumbler outlet B can be
sprayed and wiped with surface disinfectants.
Clean and disinfect the cuspidor bowl with a special cleaning agent. This
agent will also care for the drain lines of the cuspidor bowl.
Vacuum system, foaming cleaning agents
The cuspidor bowl is attached to the water unit via a bayonet catch and
can be removed for thorough cleaning. The tumbler holder A can be left
attached when doing this.
Cuspidor
Tumbler filling
NOTICE
Domestic cleaning agents foam up.
Foaming cleaning agents often cause foam and water to be sucked into
the dry suction system. This can cause damage to the suction machine.
➢ Use only care, cleaning and disinfecting agents approved by Sirona,
see"Care, cleaning, and disinfecting agents" [ → 205].
CAUTION
If the treatment center is switched on, the flushing and tumbler filling
functions can be activated even with the cuspidor bowl removed.
In this case, water would run onto the floor and could enter the treatment
center.
➢ Switch the treatment center off at the main switch before removing
the cuspidor bowl.
CAUTION
In the Compact water unit, tumbler filling can be switched on even if the
tumbler filler is attached.
In this case, water would run onto the floor and could enter the treatment
center.
➢ Do not press the
Tumbler filling
key when the tumbler filler is
removed.