2Safety instructions Sirona Dental Systems GmbH
2.8Trouble-free operation Operating Instructions INTEGO
2.8
Trouble-free operation
Use of this unit is permissible only if it works properly without
malfunctions. If trouble-free operation cannot be ensured, the unit must
be taken out of service, checked by authorized technicians for
malfunctions and, if necessary, repaired or replaced.
2.9
Vacuum system
Suction of aluminum or other metal oxides prohibited
The suction removal of aluminum and other metal oxides from blasting
devices via the amalgam separator and the automatic separator installed
in the treatment center is prohibited! This would result in extreme wear
and clogging of the vacuum and drain lines.
Separate suction
A separate vacuum system must be used in connection with metal oxide
blasting devices. Treatment centers equipped with a central standard wet
suction are generally suitable for suction removal of the above material.
However, make sure to observe the instructions provided by the
manufacturer of your vacuum system.
Suction of steel abrasives permissible
No restrictions apply when using salt blasting devices in connection with
Sirona treatment centers. However, in such cases, make sure that the
system is subsequently flushed with an adequate amount of water.
2.10
Patient chair
Maximum load capacity 140 kg or 185 kg
Please observe the maximum load capacity for the patient chair.
140 kg (308.6 lbs) forINTEGO, 185 kg (407.9 lbs) as option
185 kg (407.9 lbs) forINTEGO pro
The permissible maximum load capacity is indicated on a plate next to the
rating plate of the treatment center.
The weight distribution complies with ISO 6875. The mechanical stability
is tested with a multiple safety factor acc. to IEC 60601-1.
The maximum permissible weight of accessories mounted on the patient
chair is 5 kg (11 lbs).
The patient’s arms and legs must be resting on the upholstery of the
chair.
2.11
Intermittent operation
Intermittent operation
The motors of the treatment center and of the treatment instruments are
designed for intermittent operation corresponding to the dental mode of
treatment.
Drive motors for patient chair and backrest: max. 10% duty cycle, (max.
2 min "ON" / 18 min "OFF")
max. load of chair
185 kg
Intermittent operation
of chair motors:
2 min ON/18 min OFF
max. load of chair
140 kg
Intermittent operation
of chair motors:
2 min ON/18 min OFF
ü û
max. load of chair
185 kg
Intermittent operation
of chair motors:
2 min ON/18 min OFF
max. load of chair
140 kg
Intermittent operation
of chair motors:
2 min ON/18 min OFF