Sirona Dental Systems GmbH 2Safety instructions
Operating Instructions INTEGO 2.19Electrostatic discharge
båÖäáëÜ
SIVISION digital and USB interface
The presence of electromagnetic interference in the vicinity of the
treatment center may cause image degradation and interruptions in the
data transmission via the USB interface to the PC. In such cases, repeat
the image recording or other operations.
In the event of heavy interference, it may be necessary to restart the PC
and treatment center. It is therefore not recommended to use the PC for
controlling other devices that perform essential tasks.
2.19
Electrostatic discharge
Protective measures
Electrostatic discharge (abbreviated: ESD – ElectroStatic Discharge)
Electrostatic discharge from people can damage electronic components
when the components are touched. Damaged components usually have
to be replaced. Repairs must be performed by qualified personnel.
Measures to protect against ESD include:
Procedures to avoid electrostatic charging via
air conditioning
air humidification
conductive floor coverings
non-synthetic clothing
discharging the electrostatic charges from your own body through
contact with
a metallic unit casing
a larger metallic object
any other metal part grounded with the protective earth
Endangered regions are indicated on the unit by the ESD warning label:
We recommend that all persons working with this system are made aware
of the significance of the ESD warning label. A training course should also
be held to inform users about the physics of electrostatic charges.
Physics of electrostatic charges
An electrostatic discharge requires prior electrostatic charging.
There is a danger of electrostatic charges building up whenever two
bodies rub against each other, e.g. when:
walking (soles of shoes against the floor) or
moving (chair casters against floor).
The amount of charge depends on several factors: The charge is:
higher at low air humidity than at high air humidity, and
higher with synthetic materials than with natural materials (clothing,
floor coverings).