Switching the key sound on/off
A setting can be made to issue an acoustic signal when the operator
presses a button on the dentist or assistant element.
The setting focus
Key tone
flashes.
1. Change the value with the favorite keys 1 or 3.
If there is a 1 in the setting focus, the key tone is switched on.
2. Confirm with the
2 / Setup
favorite key.
4.14.1.5
Coupling suction to the 4-way foot switch.
If the treatment center is equipped with a position selector valve for the
suction system, it can be set so suction flow can be interrupted or
restarted by moving the 4-way foot control switch at the base of the chair
in any direction. Follow the safety instructions, see "Suction handpieces"
[ 141]
The setting focus
Couple suction at 4-way foot switch
flashes.
1. Change the value with the favorite keys 1 or 3.
If a 1 is in setting focus, the suction can be turned on/off at the 4-
way foot switch.
2. Confirm with the
2 / Setup
favorite key.
If you put the suction hose in the holder while the suction flow is
interrupted, the suction flow is automatically restarted when you pick it up
again.
4.14.1.6
Switching the water heater on/off
The optional heater for the treatment water can be switched on/off.
The
water heater
setting focus flashes.
1. Change the value with the favorite keys 1 or 3.
If there is a 1 in the setting focus, the water heater is switched on.
2. Confirm with the
2 / Setup
favorite key.
4.14.1.7
Setting the purging time for the purge function
The purging time for the purge function may be set between 20 and 120
seconds.
The setting focus
purge function
flashes.
1. Use favorite keys 1 and 3 to set the purge time.
2. Confirm with the
2 / Setup
favorite key.