Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.4Vacuum system
båÖäáëÜ
5.4.2.3
Suction hose cleaning on the Ambidextrous water unit
Suction hose cleaning with the INTEGO pro
With the INTEGO pro treatment center, the Ambidextrous water unit is
always equipped with an adapter for the suction hose cleaning function.
This can be used to purge the suction hoses consecutively. A cleaning
agent is automatically added to the water if the dental treatment center is
equipped with the chemical suction hose cleaning option.
Suction hose cleaning with the INTEGO
With the INTEGO treatment center, the suction hose cleaning for the
Ambidextrous water unit is optional. Chemical suction hose cleaning for
the INTEGO treatment center is not available.
If the INTEGO treatment center is not equipped with the suction hose
cleaning option, a large glass of cold, clear water must be regularly
aspirated instead, see "Purging the suction system" [ 235].
Cleaning suction hoses and suction system
1. Remove the suction tips from the suction hoses.
2. Set maximum suction flow on the suction handpiece to be cleaned.
All the other suction handpieces must be completely closed.
3. Attach the open suction handpiece to the receptacle.
4. Press the button.
Water and a mixture of water and cleaning agent is pumped into
the suction hose cleaning container and suctioned out of the
suction hoses. The end of the suction hose cleaning can be heard
by the suction noise. If the Ambidextrous water unit is equipped
with an amalgam separator, a signal can be heard at the end of
the suction hose cleaning.
5. Then place the suction hose back into the holder on the assistant
element.
6. Repeat this procedure for the other suction handpieces.
Vacuum system, hose fittings
CAUTION
Suction hose cleaning must not be interrupted to ensure that no residual
water remains in the container.
CAUTION
Disinfect the hose fittings on the water unit after each patient.