5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.4Vacuum system Operating Instructions INTEGO
Refill cleaning agent (for the INTEGO pro with chemical suction hose
cleaning option)
When the adjacent display appears on the EasyPad display or in the
status column of the touchscreen, the cleaning agent for cleaning the
suction system is almost all gone. It should then be refilled as soon as
possible.
Vacuum system, Mixing with DENTOSEPT P
1. Remove the maintenance flap from the water unit.
2. Take the cleaning agent tank from out of the water unit and pull out
the hose.
3. Place the cleaning agent tank onto the ground or onto a firm surface.
4. Open the lock of the cleaning agent tank and add the cleaning agent.
The cleaning agent tank has a capacity of 1 liter.
5. Lock the cleaning agent tank.
6. Reconnect the hose to the cleaning agent tank again and push the
cleaning agent tank into the water unit. The filling opening must face
the front.
7. Reattach the maintenance flap.
With the chemical suction hose cleaning option, the drain outlets of the
cuspidor are not cleaned/disinfected. If the treatment unit is equipped with
this option, the drains of the cuspidor must therefore be cleaned weekly,
see "Cleaning the drain outlets of the cuspidor" [ → 250].
Adjusting the cleaning agent mixture for chemical suction hose
cleaning
In the Setup of the treatment center, it is possible to set the amount of
cleaning agent to be mixed with water, see "Adjusting the cleaning agent
mixture for chemical suction hose cleaning" for the EasyPad [ → 195], for
the EasyTouch [ → 201].
!
Aqua
Amalg
San
1d
41d
B
10:45:03
chem
CAUTION
The cleaning agent for the vacuum system could be mistaken for the
disinfectant for the water lines.
➢ Do not fill DENTOSEPT P disinfectant into the cleaning agent tank
for chemical suction hose cleaning! Use only "Agents for the suction
lines" approved by Sirona, see Care, cleaning and disinfecting
agents" [ → 205].