4.4Patient chair Operating Instructions INTEGO
4.4.2 Safety stop
The treatment center is equipped with various safety stops to prevent
crushing and damage. The cutoff trigger points are shown in the following
illustration:
Display of triggered safety switch 2 on the EasyPad
Display of triggered safety switches (all shown in one illustration) on the
EasyTouch
The following occurs when one or more safety switches are triggered:
All safety switches cause an acoustic signal to be played twice each
time a movement stops or starts. Exception cuspidor: only when
movement starts
All chair movements stop immediately
The triggered safety switches are displayed on the EasyPad display
or on the touchscreen
As long as a safety switch is activated, the operation of the treatment
center is restricted!
If a safety switch is permanently blocked, please contact your service
technician.
2 Cuspidor bowl
3 Lift frame
4 Assistant element
5 Rear facing, right/left
6 Footrest
7 Backrest
Motor for adjusting the chair height or
backrest was switched off to avoid
overheating. The chair can be moved again
once it has cooled down.
!