5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.3Instruments and instrument hoses Operating Instructions INTEGO
5.3.3 Automatically purge water lines (autopurge function)
The autopurge function is only available on the INTEGO pro treatment
center.
The autopurge function enables automatic purging of all water-carrying
instruments in the dentist element, of the 3-way SPRAYVIT E syringe in
the dentist and assistant elements and of the tumbler filling.
All of the instruments inserted in the water unit will be purged when the
autopurge function is activated. If the instruments remain in the water unit
after the treatment center is switched off, the purging process will
automatically be started again the next time the treatment center is
switched on.
Execute the AutoPurge function:
● Before starting work
● At the end of the work day
When the treatment center is switched to stand-alone water supply mode,
the AutoPurge function is no longer available (function hidden). It is still
possible to purge individual instruments; see "Purging water lines (purge
function)" [ → 214].
Preparation
The following preparations must be made before you start purging the
water lines:
1. If your treatment center is equipped with a cuspidor, activate the
cuspidor flushing for at least one minute. This flushes the water lines.
2. Set all burr drives and the SIROSONIC L scaler to the maximum
water flow rate. The water regulators under the instrument holder
must be turned counterclockwise.
3. Put all instruments in place.
4. Do not remove the tumbler holder from the cuspidor. Place the empty
tumbler under the tumbler outlet.
5.3.3.1
Autopurge function on the EasyPad
Setting the autopurge time
The purge time for automatic purging may be set between 60 and 180
seconds in the treatment center setup, see "Setting the purging time for
the autopurge function" [ → 195].
Switching the autopurge function on
✔ All instruments are in their holders.
IMPORTANT
The water flow rate to the instruments is not checked by the treatment
center.
➢ Make sure that the maximum water flow rate is set for all
instruments to be purged.