Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.3Instruments and instrument hoses
båÖäáëÜ
The water line purging procedure is finished. The treatment center is
again ready for operation.
Interrupting the purging process
➢ Press the foot pedal to interrupt the purging process. If the instrument
is not put back in the holder, the purging process can be continued by
pressing the foot pedal again up to the end of the set purging time.
Canceling the purge function
In the case of the error message
Deposit instruments
or during purging,
the purge function can be canceled.
➢ Touch the
Stop
key on the touchscreen.