5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.6Sanitizing Operating Instructions INTEGO
1. Hold the 3-way syringe over the cuspidor or over a watertight
container with sufficient capacity and press the water key for at least
15 seconds.
2. Remove a single instrument from holder and hold it over the cuspidor
or over a watertight container with sufficient capacity. For the CS
dentist element, move the swivel arm of the instrument into the
working position.
3. Step on the pedal of the foot control briefly.
The removed instrument is purged with water for approx. 15
seconds. The purging process stops automatically.
4. Afterward, return the instrument to the instrument holder.
5. Repeat the procedure for all other instruments.
6. After you have flushed the last instrument, press the orange flashing
Stop
key.
The sanitation process is finished. The treatment center is again
ready for operation.
7. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
Interrupting instrument purging
Step on the foot pedal to interrupt the instrument purging process. If
the instrument is not put back in the holder, the purging process can
be continued by pressing the foot pedal again.
Canceling the sanitizing process
The sanitizing process can be canceled, e.g., if the treatment center must
be switched on again before 24 hours have passed.
The sanitizing process has started.
Touch the
Stop
key on the touchscreen.
The sanitation process is canceled. If the water lines have
already been treated with DENTOSEPT P disinfectant, they have
to be rinsed with water before use. The display
Refill water
appears, see "Filling the fresh water bottle with water" (above).
Cleaning the suction hoses
After sanitizing with the fresh water bottle, the suction hoses should also
be chemically cleaned, see "Cleaning suction system using cleaning
adapter in the cuspidor or via external container" [ 242].