Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.6Sanitizing
båÖäáëÜ
2. Remove a single instrument from holder and hold it over the cuspidor
or over a watertight container with sufficient capacity. For the CS
dentist element, move the swivel arm of the instrument into the
working position.
3. Step on the pedal of the foot control briefly.
The removed instrument is then purged for approx. 15 seconds
with DENTOSEPT P disinfectant. The purging process stops
automatically.
4. Afterward, return the instrument to the instrument holder.
5. Repeat the procedure for all other instruments.
Tumbler filling, discolorations due to DENTOSEPT P
Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
1. Do not press the orange flashing
Stop
key.
2. Switch the treatment center off via the mains switch A on the base of
the chair. Make sure that the treatment center is switched off for at
least 24 hours, but not longer than 3 days (maximum sanitation time).
3. Empty the rinsing tumbler and return the empty tumbler below the
tumbler outlet.
Switch the treatment center on again after 24 hours and before 3
days have elapsed.
Filling the fresh water bottle with water
The display
Refill water
appears on the touchscreen.
1. Unscrew the fresh water bottle counterclockwise from the holder on
the water unit and empty the fresh water bottle completely.
2. Mix food-grade water (< 100 CFU/ml) with the DENTOSEPT P
disinfectant at a ratio of 100:1 (1 liter of water, 10 ml DENTOSEPT P)
and fill this into the fresh water bottle. The bottle has a capacity of
approx. 1.3 liters.
3. Attach the hose to the fresh water bottle and screw the bottle
clockwise into the holder.
4. Touch the
OK
key.
The
Stop
key flashes orange on the touchscreen. Tumbler filling
is switched on and is flushed with water.
Purging water lines with water
The DENTOSEPT P disinfectant is rinsed out of the instrument hoses,
the SPRAYVIT hoses, and tumbler filler with water.
The tumbler filling was automatically flushed with water.
max.
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist
cloth.