EasyManua.ls Logo

Sirona Intego - Page 243

Sirona Intego
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 57 498 D3543
D3543.201.01.05.02 08.2016
243
Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.4Vacuum system
båÖäáëÜ
5. Remove the suction tips from the suction hoses.
6. Set maximum suction flow on the suction handpieces.
7. Attach the adapter C to the saliva ejector.
Cleaning procedure
1. Pour 1 liter of cleaning solution into the cuspidor.
2. Remove the suction hoses from their holders and attach them to the
side of the cleaning adapter as simultaneously as possible.
Two thirds of the cleaning solution will be suctioned off by the
suction hoses; one third will flow into the drain of the cuspidor.
3. Allow the cleaning solution to act. Observe the reaction time specified
for the cleaning solution by the manufacturer.

Table of Contents

Other manuals for Sirona Intego

Related product manuals