5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.5Components of the water unit Operating Instructions INTEGO
1. With the Compact or Comfort water unit: Open the flap A on the base
of the water unit.
2. Loosen the bayonet catch by turning the lower part of the amalgam
separator B counterclockwise. Remove the lower part of the
amalgam separator along with the amalgam rotor located inside it.
The message "Amalg" appears on the touchscreen and an
acoustic signal sounds.
3. Properly dispose of the amalgam residues collected from the
cuspidor together and from the suction hoses in the amalgam rotor,
see "Cleaning the gold trap" [ 248] and "Emptying the central
suction sieve" [ 246]. Fill the amalgam residues into the amalgam
rotor.
4. Hold the lower part of the amalgam separator upright. Attach the
transport cap to the amalgam rotor.
The transport cap locks in place. A bonding agent is released on
attachment. Therefore, please note that the transport cap cannot
be removed again once it has been closed.
5. Remove the amalgam rotor with the transport cap from the lower part
of the amalgam separator.
6. Place the container in the special packaging and ship it for disposal
or authorize a certified waste management company.
Installing the amalgam rotor
Use only original Sirona accessories. Never use a used or recycled
amalgam rotor.
To reorder the amalgam rotor, see "Spare parts and consumables"
[ 299].
1. Grease the O-ring E on the lower part of the amalgam separator.
Lubricants, see "Care, cleaning, and disinfecting agents." [ 205]
2. Insert the new amalgam rotor C in the lower part of the amalgam
separator B.