Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.5Components of the water unit
båÖäáëÜ
1. Open the flap A on the base of the water unit.
2. Loosen the bayonet catch by turning the sediment container B
counterclockwise.
3. Pour the excess water out of the sediment container and collect the
amalgam residues. Properly dispose of the amalgam residues
together with the amalgam residues collected from the cuspidor and
from the suction hoses, see "Cleaning the gold trap" [ → 248] and
"Emptying the central suction sieve" [ → 246]. Authorize a certified
waste management company for this purpose.
Installing the sediment container
1. Grease the O-ring C on the sediment container. Lubricants, see
"Care, cleaning, and disinfecting agents" [ → 205].