Sirona Dental Systems GmbH 5Care, cleaning and maintenance by the practice team
Operating Instructions INTEGO 5.5Components of the water unit
båÖäáëÜ
3. Pour the excess water out of the container and collect the amalgam
residues. Properly dispose of the amalgam residues together with the
amalgam residues collected from the cuspidor and from the central
suction sieve, see Cleaning the gold trap" [ → 248] and Emptying the
central suction sieve" [ → 246]. Authorize a certified waste
management company for this purpose.
Installing the collector of the air jet pump
1. Hold the sediment container B so that the latching noses of the slide
lock are positioned transverse to the water unit. Rotate the sediment
container B clockwise.
The collector is locked in place.
2. Close the flap A.
3. Flush the cuspidor for at least one minute.