Sirona Dental Systems GmbH 7Malfunctions
Operating Instructions INTEGO 7.3Error Messages
båÖäáëÜ
The error message or warning triangle automatically disappears when the
error has been remedied. If it does not disappear, please inform your
service technician.
27 Suction hose cleaning Too little water is used for suction
hose cleaning.
Inform your service technician.
Instead, aspirate a glass of water
regularly, see “Rinsing the suction
system” [ → 235].
28 Chair motor If a drive motor of the patient chair
heats up excessively, the chair will
not move.
Wait until the motor cools down.
Please observe the maximum load
capacity of the chair, see “Patient
chair”.
Inform your service technician if the
error occurs repeatedly.
30 Central data storage The treatment center was reset to
factory settings. Settings are not
saved.
Inform your service technician.
31 Chemical suction hose
cleaning
The cleaning agent tank for chemical
suction hose cleaning has not been
emptied since the last filling despite
several suction hose cleanings.
Inform your service technician.
32 Motor configuration Configuration of the motor control
does not correspond with the
connected motor. The motor does not
operate at the set speed or comes to
a standstill.
Connect the BL E motor to the
instrument hose instead of the BL ISO
E or vice versa.
The motor control can be set to the
motor version used in the service
area of the setup dialog. Inform your
service technician.
33 Cuspidor bowl valve The water container of the cuspidor
valve is not emptied. The tumbler
filling, flushing, suction hose cleaning,
sanitation, purge and autopurge
functions cannot be activated.
Inform your service technician.
Code Error Description Measure