4Operation Sirona Dental Systems GmbH
4.5Dentist element Operating Instructions INTEGO
4.5.7.2.1 Setting the highspeed handpiece lamp on the EasyPad
Accessing the turbine light operating context
1. Detach the turbine to be set from the holder.
2. Press and hold the function key
2 / Setup
(> 2 s).
The
Instrument light
operating context is displayed on the
EasyPad screen. The selection focus
Voltage
(U) flashes.
Limiting voltage
✔ The selection focus
Voltage
(U) flashes.
1. Press the function key
2 / Setup
.
The setting focus
Setting the voltage
flashes.
2. Set the voltage for the lamp of the turbine selected in volts using
favorite keys 1 and 3.
3. Confirm the setting by pressing the function key
2 / Setup
.
The voltage is limited to the level set. The operating context
Turbine
appears.
Limiting current
✔ The selection focus
Voltage
(U) flashes.
1. Press favorite key 3.
The selection focus flashes
Current strength
A flashes.
2. Press the function key
2 / Setup
.
The setting focus
Setting the current strength
flashes.