5.17 Castors Maintenance
To replace any of the castor mechanisms, you must tip the
BleaseSirius. This positioning requires AT LEAST 2 people to
perform.
Remove the Vaporizers, monitoring, cylinders, all loose items
and all add-ons from the machine.
With the assistance of at least one additional person, tip
BleaseSirius onto its back, preferably on a protective surface.
Be sure to prevent the rear castors from rolling as you lay it
down. Lay the BleaseSirius on its back to access all 4 castors.
Figure 130 - Caster Removal
Figure 129 - Location of Caster Retaining
Screw
1. Remove the castor retaining
screw.
2. Remove the castor.