200
5.29 Reprogramming
The software for the Blease700/900 series of ventilator consists of two parts,
the control software and the monitor software.
The control software is held on the BAV controller (FR101027). A replacement
BAV controller will have the latest version of software pre-installed.
The monitor software is held on the display board (FR101026). A replacement
display board will have the latest version of monitor software pre-installed.
The version of software installed on the ventilator can be seen on the screen
by selecting Setup, Confi guration, System Information. The display software
will be displayed next to the FRONT PANEL software and will be in the format
V700900 x.ox followed by the date. The control software will be displayed next
to control board software and will be in the format BAV040-700900-V9_5x_
followed by the date.
There are three occasions when the software needs to be upgraded:
To update the software version
To update the language option
To update the ventilator model
5.29.1 Software Upgrade Procedure
Three fi les are required for software upgrade.
Ventfl ash 2.1.0.2.exe This is a download program.
The program fi le for Control Processor, BAV040_700900_V10_5x.ABS
The program fi le for the display processor,
FP700900_9_1X_combined_A20
Detailed Repair Procedures

Table of Contents