4.16 Vaporizer Output Concentration Check
1. Verify that the vaporizer being tested is fi lled to at least half its capacity.
2. Set the vaporizer dial to “0”.
3. Calibrate a gas analyzer (i.e. Riken) according to the manufacturer’s
instructions.
4. Ensure a waste gas scavenging systems is on, operational and is
connected to the absorber.
5. Connect a 22 mm reusable tube to the inspiratory and expiratory ports on
the absorber. Connect a breathing bag to the bag port.