262
19. The following equipment is mandatory, please refer to the relevant user
manuals for instructions in usage; blood oxygen monitor, CO
2
alarm,
blood pressure alarm, ECG alarm, anesthesia agent concentration.
20. The machine is to be equipped with an anesthetic gas scavenging
transfer and receiving system, complying with ISO 8835-3 or
applicable local standard before being put into service.
9.9.2 Electrostatic Sensitive Devices (ESD) Warnings and
Cautions
Notices and Important Information
All ESD must be stored in approved conductive packaging, tubes,
shipping bags, foam or tote bins.
All persons handling ESD must be properly grounded via a 1MΩ
resistive grounded wrist strap.
Cover all ESD bench tops with grounded conductive mats and
connect all work surfaces and equipment to earth ground.
Transport all assemblies containing ESD in a conductive bag or
container.
DO NOT use cellophane adhesive tape to wrap DIP (dual inline
package) tubes together.
DO NOT handle ESD by their pins or mix them with other routine
electronic parts.
Never place ESD on ungrounded surfaces or leave them unattended
in an open area.
Avoid cellophane wrappers, synthetic (non conductive) carpeting,
warm or cool air blasts, Styrofoam coffee cups, etc. when working
with ESD.
Use only properly designed heat lamps, heat chambers and/or
‘antistatic’ quickchill sprays during troubleshooting or stress testing
procedures.
NOTE: In particular electronic assemblies in the BleaseSirius®
range of machines are easily damaged by ESD and require
special handling.