56
Planned Maintenance
3.7 CLEANING
Cleaning of external surfaces is possible using water or isopropyl alcohol.
Cleaning of internal surfaces and parts should not be required and is not
recommended.
This product is intended for use in medical compressed gas
systems only. Do not use this product where pressures and
temperatures can exceed those listed under Technical Data.
DO NOT use any lubricants that are not compatible with oxy-
gen or medical gases. The use of lubricants not compatible
with oxygen may pose a risk of fi re or explosion.
Open the cylinder valve slowly to prevent the risk of re or explosion due to
oxygen pressure shocks. The relief valve is a safety device and must only be
removed for servicing. When servicing is complete, a new relief valve must be
tted. Do not attempt to adjust or tamper with the relief valve.