System Checks
4.1 Overview
This chapter explains the systems checks that must be used on all BleaseSirius
anesthetic machines whenever any of the following procedures has been
performed:
• Installation
• Planned Maintenance (should be performed annually)
• Service
• Repair
Use the service check sheet, shown on page 64, to record the results.
4.2 Tools and Test Equipment Required
1. 2 mm, 2.5 mm, 3 mm, 4 mm, 1/8 inch Allen key
2. 13 mm wrench
3. Adjustable crescent wrench
4. Pressure measuring device (able to read 0 – 100 psi)
5. Pressure measuring device (able to read 0 – 100 cmH
2
O)
6. Flow measuring device (able to read 0 – 100 L/min)
7. Flow measuring device (able to read 20 – 1000 ml)
8. Oxygen analyzer
9. Pressure Gauge (able to read 0 – 200 mmHg)
10. Copy of the BleaseSirius Checkout Sheet (See page 64). You can use a
print copy (073-0301-00) or an electronic copy (073-0321-00).
Contact Technical Support at
advsupport@spacelabs.com to request an
electronic copy of the BleaseSirius Checkout Sheet.