EasyManua.ls Logo

Spacelabs BleaseSirius - Figure 41 - Connect a Pressure Measuring Device

Spacelabs BleaseSirius
298 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
86
System Checks
22. Check that the inspired volume (VTI) on the fl ow measuring device is
500 ±50 ml.
23. Increase the tidal volume to 750 ml. Wait for the ventilator to stabilize.
24. Check that the inspired volume (VTI) on the fl ow measuring device is
750 ±75 ml.
If steps 20, 22 or 24 should fail, perform a PRAAA calibration ac-
cording to the instructions given in PSN: 076-9027-00.
25. Place the ventilator in Standby. Connect a pressure measuring device to
the inspiratory port of the absorber (Figure 41).
Figure 41 - Connect a Pressure Measuring Device
26. Set the bag/vent switch to Vent.
27. Ensure that the bellows is full.
28. Set the PEEP on the ventilator to 10 cmH
2
O.
29. Verify that the pressure measuring device reads 10 ±1.5 cmH
2
O.
Change the ventilator from volume control to pressure control mode at
the following settings: Pressure = 10 cmH
2
O
BPM = 8
I:E = 1:2
PEEP = OFF
30. After 12 breaths, verify that the pressure measuring device reads
10 ± 2 cmH
2
O.

Table of Contents