EasyManua.ls Logo

Spacelabs BleaseSirius - Figure 18 - Connect Pressure Measuring Device; Figure 19 - Regulator Nut and Set Crew Locations

Spacelabs BleaseSirius
298 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
System Checks
4. Disconnect the 6 mm plug from the regulator test point, then connect a
pressure measuring device to the test point.
5. Turn the cylinder on.
6. Turn the switch to the “ON” position and set a fl ow of 1 L/min on the fl ow
control valve for the appropriate gas being tested.
7. Verify that the pressure measuring device reads between 45 ± 2 psi.
8. If the reading is out of spec, loosen the nut on the regulator being tested
using a 13 mm wrench. Adjust the set screw using a 1/8 inch Allen key to
set the regulator to 45 ± 0.5 psi. Once achieved, secure the nut.
9. Repeat steps 3-8 for all regulators. (Note: When testing outputs or making
adjustments on N
2
O regulators, ensure that an O
2
supply is connected to
the O
2
pipeline inlet.)
Figure 18 - Connect Pressure
Measuring Device
Figure 19 - Regulator Nut and Set Crew Locations

Table of Contents