EasyManua.ls Logo

Spacelabs BleaseSirius - 1.4 Technical;Performance Specification

Spacelabs BleaseSirius
298 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
1.4 Technical/Performance Specifi cation
1.4.1 Controls
Oxygen fl ow 150ml/m to 10 l/m Simplex/ Cascade
Nitrous oxide fl ow 0ml/m to 12 l/m Simplex/Cascade
MED AIR fl ow 0ml/m to 15 l/m
Flowblock assembly accuracy ±5% measured value at 20ºC and
101.3 kPa/14.6 psi
Oxygen fl ush Non-locking 35 to 55 l/m
Vaporizers Accepts Selectatec
Hypoxic gases Minimum 21% oxygen/nitrous oxide
mixture allowed
1.4.2 Ventilator
A Blease700/900 ventilator is built
into the BleaseSirius.
1.4.3 Alarms / Indicators
Oxygen failure Audible alarm sounds for minimum
of 8 secs when oxygen pressure falls
below 30 psi.
1.4.4 Regulator Safety Valve Settings
Cylinder regulator 43 psi - 47 psi
Cylinder regulator relief valve > 75 psi
Machine gas piping design
rating
700 kPa/max. 101.5 psi/max
Secondary hypoxic regulators
25-32 psi
25-35 psi
N
2
O
}
O
2
0.5 Lpm fl ow
Common gas outlet relief valve 3.25 psi - 3.95 psi
Technical Description

Table of Contents