1.4 Technical/Performance Specifi cation
1.4.1 Controls
Oxygen fl ow 150ml/m to 10 l/m Simplex/ Cascade
Nitrous oxide fl ow 0ml/m to 12 l/m Simplex/Cascade
MED AIR fl ow 0ml/m to 15 l/m
Flowblock assembly accuracy ±5% measured value at 20ºC and
101.3 kPa/14.6 psi
Oxygen fl ush Non-locking 35 to 55 l/m
Vaporizers Accepts Selectatec
Hypoxic gases Minimum 21% oxygen/nitrous oxide
mixture allowed
1.4.2 Ventilator
A Blease700/900 ventilator is built
into the BleaseSirius.
1.4.3 Alarms / Indicators
Oxygen failure Audible alarm sounds for minimum
of 8 secs when oxygen pressure falls
below 30 psi.
1.4.4 Regulator Safety Valve Settings
Cylinder regulator 43 psi - 47 psi
Cylinder regulator relief valve > 75 psi
Machine gas piping design
rating
700 kPa/max. 101.5 psi/max
Secondary hypoxic regulators
25-32 psi
25-35 psi
N
2
O
}
O
2
0.5 Lpm fl ow
Common gas outlet relief valve 3.25 psi - 3.95 psi