System Checks
Figure 20 - Test Hose Assembly
4.9 High Pressure Outlet Test
4.9.1 Parts required
Only order parts that are required for testing, depending on which high pressure
outlet is fi tted to the machine.
O2 Schraeder (White) Air Schraeder (Black) O2 DISS (Green)
10700007, Quantity: 1 10700008, Quantity: 1 10700009, Quantity: 1
376-0559-00, Quantity: 1 376-0559-00, Quantity: 1 376-0559-00, Quantity: 1
ST2161, Quantity: 1 ST2161, Quantity: 1 ST2161, Quantity: 1
ST4710, Quantity: 1 ST4710, Quantity: 1 ST4710, Quantity: 1
4.9.2 Test Hose Assembly
1. For O
2
and Air Shraeder, cut the NIST tting end of the hose. For O
2
DISS, cut either end of the hose.
2. Fit the cut end of the hose through the O-clip.
3. Fit the 22 mm female connector through the hose and secure the
assembly with the O-clip.