5.19.9 Removal of Flow Control Valves (up to March 2009)
From March 2009 the Flow Control Valve and Gas Inlet Module have
been combined into one unit (FR137027). The valves must be replaced
as a pair only. See 137P1004 in Appendix 1 for full replacement and
test details.
1. Disconnect the four
spade connectors and
the hoses as shown.
2. Remove the two nuts as shown, which in turn releases the valves.