Ventilator Passwords
FAN FAIL Select this option to turn the fan fail alarm
on and off. For stand alone versions of the
ventilator, this option will remain unchecked.
For integrated ventilators this will be checked.
Absorber Bypass A Spacelabs Healthcare future option.
cmH20/hPa Select this option to change displayed
pressure units from cmH20 to hPa.
Touch Sensitivity Select this option to allow the Display to return
to home screen and allow the user to verify
the accuracy of the touchscreen calibration
by leaving a trace of all active touches. The
unit must be powered down to exit.
CALBK Service Menu
Select: Setup>Confi guration>Service Mode>CALBK
FRU Defaults FRU Defaults is required when the Inspiratory
Block, the BAV controller, or Front Panel is
replaced. Select this option to set defaults on
component being replaced.