EFM Option
6.1.2 Technical Specifi cations
Controls O
2
ow: 150 ml/m to 15 lpm
N
2
O fl ow: 0 ml/m to 12 lpm
Med Air fl ow: 0 ml/m to 15 lpm
Graphics Resolution Log spaced from 0.1 lpm to 2 lpm
Lin from 2 lpm to 13 lpm
Flowblock Accuracy* O
2
: 10% - 100% Full scale ± 10% of reading or
150ml/min when <10% of full scale.
N
2
O: 10% - 100% Full scale ± 10% of reading or
150ml/min when <10% of full scale.
AIR: 10% - 100% Full scale ± 10% of reading or
150ml/min when <10% of full scale.
Total gas fl ow: ± 8% of reading
*Accuracy at 21°C (70°F) and 1013mbar
Hypoxic Gases Minimum oxygen concentration of 21% for oxygen/
nitrous mixtures
Alarms Unrecoverable errors are displayed on the
EFM LCD by the warning triangle.
Three errors are identifi ed on the ventilator screen
by the alarm message, CHECK TOTAL FLOW. The
three conditions triggering this alarm are:
Oxygen fl ow is less than 100 ml/min.
Gas fl ow is absent (indicated on the EFM display
by dashed lines).
The ventilator receives no fl ow data from the
EFM for more than 45 seconds.