269
port of the bellows base. This will increase the pressure inside the bellows
and cause it to detach from the base, resulting in a serious malfunction. Any
problem arising from an improperly functioning scavenging system is solely the
responsibility of the user.
8. DAMAGE TO INTERNAL COMPONENTS
Excessive penetration of fi xing screws into the ventilator may damage internal
components. Ensure that the screws do not penetrate more than 7mm/0.2” into
the ventilator.
9. CLEANING AGENTS
The ventilator surfaces are not scratch-resistant. Do not use abrasive cleaning
agents otherwise damage to the surfaces will result.
Chemical decontaminants or liquid sterilization agents will damage the sensor
and must NOT be used for cleaning or sterilizing. If autoclaving the patient
sensor, the autoclave must only be used with distilled water.
10. VALVE SEAT
If the bellows valve seat is damaged, the pop-off valve will leak and may cause
serious malfunction. Take care not to damage the precision-molded surface of
the valve seat while cleaning. Never use a hard object or abrasive detergent.
Use only a soft, lint-free cloth.
11. STERILIZATION (bellows)
To avoid damage to the equipment:
• Peak sterilization temperature must not exceed 134ºC.
• Do not sterilize the control unit.
• Gas sterilization should be followed by quarantine in a well-ventilated
area to allow dissipation of residual absorbed gas.
• Follow the sterilization agent manufacturer’s instructions.
12. CALIBRATION PORT
Improper connection of equipment or any power sources to the calibration port
may permanently damage the ventilator. Only a qualifi ed technician should
connect monitoring equipment to this port. Such equipment must be Spacelabs
Healthcare -compatible and meet the regulatory standards of the countries in
which the ventilator is used.
13. FLOW SENSOR TUBING
Check fl ow sensor tubing regularly for trapped water as it could affect ventilator
readings.
Notices and Important Information