USTER
®
HVI SPECTRUM
Instruction Manual
Uster Technologies, Inc. Chapter 14
Appendix • Page 14-25
“Receive data register full
in Control Board”
The receive data register filled up
without being processed in the
control board.
“Receive buffer overflow in
Control Board”
The receive buffer overflowed in the
control board.
“Noise error in Control
Board”
A value that was not valid data was
received in the control board.
“Framing error in Control
Board”
The byte was not properly framed
when received by the control board.
“ERROR: LOW AMOUNT ->”
During Testing, the sample of cotton
taken was too small to give good
strength results. The Length/Strength
cycle must be repeated.
“ERROR: HIGH AMOUNT ->”
During Testing, the sample of cotton
taken was too large to give good
strength results. The Length/Strength
cycle must be repeated.
“Read Video Configuration
File failed”
The Video configuration file (vp.cfg)
could not be read.
“Xilinx Download failed”
The Framegrabber card could not be
programmed.
“The number of color/trash
tests must be 1 or 2 times
the number of
length/strength tests”
On the SETUP, CONFIGURATION
page, under the MEASUREMENTS TAB,
the number of color/trash tests must
be either the same as the number of
length/strength tests, or exactly twice
the number.
“Error: Positioner not at back
of middle sensor before
starting Length/Strength
cycle”
Reset and initialize Length/Strength
unit to continue.