FUNCTIONALITY
GUI - Serial numbers for transformers no longer auto increment on stop or safety
stop. They only increment on a result of "PASSED" or "FAILED".
GUI - Safety stops are now only posted if a program, self-test or compensation is
running.
GUI - Added a box around text in the execution screens around instruction text. This
differentiates the instructions from the status information.
GUI - Changed the button "CANCEL" to "SUSPEND" on the results return error
screen.
GUI - Changed the button "CANCEL" to "SUSPEND" on the printer error screen.
GUI - Increased the size of the top two touch areas in the part list to decrease the
likelihood of a missed touch.
GUI - Returning to the program execution screen after a user or safety abort on the
first run of a program no longer corrupts the Custom Text field. The proper stop
screen is shown instead.
GUI - The results screen no longer jumps when at the bottom page after a full screen
paint
COMMS - Fixed communications with the server locking the product.
COMMS - Server failures no longer cause a problem if they happen at the same time
as a printer failure.
COMMS - The AT no longer freezes when the printer runs out of paper. The printer
error screen is shown instead, and the user can retry the print operation.
COMMS - Programs can now be run after a SUSPEND COMMS to server if user
chooses.
COMMS - The pass LED + REMOTE signal is now illuminated after power on offsets
and self-test is finished. This matches AT3600 behavior and helps with customers’
existing integration systems
TESTS
TRL - Fixed accuracy and 0020 error code in TRL.
R - Improved the accuracy of drift correction for resistance test over unit warm up
period.
STRW - STRW now has extra stabilization before trimming and measure.