6166 DC Voltage Current Source Operation Manual
4.2.13 MENU Key (Parameter Setting)
4-5
Polar Type Selects either bipolar or unipolar operation.
Bipolar: The polarity switches across zero in changing the source value with the
rotary knob.
Unipolar: The polarity does not change, stopping at zero in changing the source
value with the rotary knob.
Discharge SW Sets the discharge switch.
OFF: Discharge switch OFF
ON: Discharge switch ON
For more information, refer to Section 5.1.4, "Selecting Standby Status."
MEMORY Sets the program memory-related parameters used in the program function.
Memory Save Saves the setting program memory data to the non-volatile memory.
The saved data is loaded at power ON.
Memory Load Loads the program memory data saved in the non-volatile memory.
Memory Clear Clears (initializes) the setting program memory data.
The data saved in the non-volatile memory is not initialized.
TEMP Sets the thermal electromotive force output -related parameters.
TC Out Sets the thermal electromotive force output to ON or OFF.
OFF: Sets the thermal electromotive force output to OFF.
ON: Sets the thermal electromotive force output to ON.
TC Types Selects the type of thermocouple.
Type T: Selects type T (CC).
Type J: Selects type J (IC).
Type E: Selects type E (CRC).
Type K: Selects type K (CA).
Type S: Selects type K (PR10).
Type R: Selects type R (PR13).
Type B: Selects type B (PR30).
Type N: Selects type N.
RJC Temp Sets the reference junction temperature.
The setting range is from -270C to +120C.
Note that the reference junction temperature less than 0C is deemed to be 0C for
a type B thermocouple
JIS Standard Selects the standard JIS C1602-1995 or JIS C1602-1981.
Note that JIS C1602-1995 is applied for a type N thermocouple even if C1602-1981
is selected.
C1602-1995: Selects JIS C1602-1995 as standard.
C1602-1981: Selects JIS C1602-1981 as standard.