EasyManuals Logo

ADCMT 6166 User Manual

ADCMT 6166
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
6166 DC Voltage Current Source Operation Manual
4.2.13 MENU Key (Parameter Setting)
4-6
PARAM Loads or saves the setting parameters.
Power ON Load Selects the setting parameter condition at power ON.
Power OFF: Start up using the parameters saved when the power was last turned
off.
USER0: Start up using the parameters saved in the non-volatile memory,
USER0.
Param Save Saves the setting parameters to the non-volatile memory.
USER0: Saves the current setting parameters to the non-volatile memory, USER0.
USER1: Saves the current setting parameters to the non-volatile memory, USER1.
USER2: Saves the current setting parameters to the non-volatile memory, USER2.
USER3: Saves the current setting parameters to the non-volatile memory, USER3.
USER0-3 Default: Saves the factory default values to the non-volatile memory,
USER0 to USER3
Param Load Loads the setting parameters saved in the non-volatile memory.
Not displayed on the menu screen while Operate is ON.
USER0: Loads the parameters from the non-volatile memory, USER0.
USER1: Loads the parameters from the non-volatile memory, USER1.
USER2: Loads the parameters from the non-volatile memory, USER2.
USER3: Loads the parameters from the non-volatile memory, USER3.
Default: Loads the factory default values as setting parameters.
Initialize Initializes all the setting parameters to the factory default values.
I/F Selects or sets the interface.
Bus Selects the interface.
GPIB: Selects the GPIB interface.
USB: Selects the USB interface.
BCD: Selects the BCD interface.
BCD is displayed only in the 6161-compatible operation.
Address Sets the GPIB address from 0 to 30.
NOT displayed when the USB interface is selected.
Header Sets the header of the talker format to ON or OFF.
Displayed only in the 6161-compatible operation.
OFF: Not output the header.
ON: Outputs the header.
Listen Only Switches between Addressable and Listen only.
NOT displayed when the USB interface is selected.
OFF: Addressable
ON: Listen only
Id Sets the USB ID from 1 to 127.
NOT displayed when the GPIB interface is selected.