6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-43
Class Item Command Description
Default Operation
Power
ON
Factory
shipme
nt
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL
Cali-
bra-
tion
Calibration
range
XRn Calibration range setting
----
XR? Query of calibration range
----
Response: XR1 to XR7, XR9
Calibration
execution
XMD0 Normal source mode *Unnecessary for con-
nection coefficient calibration
----
XMD1 DMM input mode (XDT command enabled)
XMD? Normal source mode/Query of DMM input
mode
----
Response:XM0 or XM1
XNT0 Zero point calibration (An error occurs in
connection coefficient calibration.)
----
XNT1 + Full scale calibration (An error occurs in
connection coefficient calibration.)
XNT2 - Full scale calibration (An error occurs in
connection coefficient calibration or limit
calibration.)
XNT? Query of calibration point zero/full scale set-
ting
----
Response: XNT0 to XNT2
Connection
coefficient
calibration
XK0 KTU0 Upper digit output calibration
----XK1 KTU1 Lower digit output calibration
XK2 KTD Upper/lower digit output calibration
XK? Query of connection coefficient calibration
----
Response: XK0 to XK2
Calibration
data input
XDT data: DMM reading data input
(If the DMM input value is within the nor-
mal range after this command is executed,
the mode is automatically released.)
----
Voltage range Current range
XR1 - 1 mA range
XR2 10 mV range 10 mA range
XR3 100 mV range 100 mA range
(Current limit)
XR4 1 V range -
XR5 10 V range
(Connection
coefficient cali-
bration)
-
XR6 100 V range -
XR7 1000 V range
(Voltage limit)
-
XR9 1000 mV range -