6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-42
System Parameter load *RCL0 Loads data from the non-volatile mem-
ory, USER0 as setting parameters.
--
*RCL1 Loads data from the non-volatile mem-
ory, USER1 as setting parameters.
*RCL2 Loads data from the non-volatile mem-
ory, USER2 as setting parameters.
*RCL3 Loads data from the non-volatile mem-
ory, USER3 as setting parameters.
RINI Loads the factory default settings as set-
ting parameters.
--
Maximum
relay count
ORLY? Reads the maximum relay count.
Response:
dddddddddd (10-digit integers)
Cali-
bration
Calibration
mode
CAL0 OFF (Exits the calibration mode.)
CAL1 ON (Enters the calibration mode.)
CAL? Query of calibration mode
Response: CAL0 or CAL1
Calibration
data
XINI Initializes the calibration data area.
- Not initialize calibration data in the
non-volatile memory.
----
XCLR Initializes the present calibration data
such as function, range and polarity.
- Not initialize calibration data in the
non-volatile memory.
----
XWR Saves calibration data to the non-vola-
tile memory.
----
XOUT Cancels the calibration mode (OFF).
----
The calibration coefficient is not saved.
Calibration
function
XF5 Connection coefficient calibration
----
XF6 Connection coefficient calibration
XF0 Voltage source calibration
XF1 Divider voltage source calibration
XF2 Current source calibration
XF3 Voltage limit calibration
XF4 Current limit calibration
XF? Query of calibration function
----
Response: XF0 to XF6
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL