6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-41
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL
System Limiter detec-
tion buzzer
UZ0 Limit detection buzzer OFF
--
UZ1 Limit detection buzzer ON
UZ? Response: UZ0 or UZ1
-
Display bright-
ness
DBRd d: 1 (low) to 8 (high)
6--
DBR? Response: DBR1 to DBR8
LED bright-
ness
LBRd d: 1 (low) to 3 (high)
2--
LBR? Response: LBR1 to LBR3
Screen OFF DM0 Screen ON
--
DM1 Screen OFF
DM2 Screen OFF only in remote control
DM? Response DM0 to DM2
Display line DH0 1-line display
--
DH1 2-line display
DH? Response: DH0 or DH1
Interlock INL0 Interlock disabled
--
INL1 Interlock enabled
INL? Response: INL0 or INL1
Option *OPT? Query of option
Response: "0"
"
Compliance Voltage 1200 V
"
"BCD I/F"
Para Parameter
save
*SAV0 Saves setting parameters to the non-vol-
atile memory, USER0.
--
*SAV1 Saves setting parameters to the non-vol-
atile memory, USER1.
*SAV2 Saves setting parameters to the non-vol-
atile memory, USER2.
*SAV3 Saves setting parameters to the non-vol-
atile memory, USER3.
SINI Sets the factory default settings to
USER0 through USER3.
--