Page #208 background imageLoading...
Page #208 background image
6166 DC Voltage Current Source Operation Manual
6.6.4 Remote Command List
6-40
Remot
e
Operation
complete
*OPC After all operations are complete, sets
LSB of Standard Event Status Register.
*OPC? Response: 1 (after all operations com-
pleted)
*WAI Waits until all operations are complete
(GPIB only).
CMP0 6161-compatible mode OFF
CMP1 6161-compatible mode ON
Not available for the USB interface
CMP? Query of 6161-compatible mode
Response: CMP0 to CMP1
System Initialization C Device clear
*RST Parameter initialization
Instrument
information
*IDN? Tells the instrument to ask itself.
Response:
ADC Corp.6166,nnnnnnnnn,xxxx
Self test *TST? Executes the test and reads out the
results.
--
Response: 0 Pass
1 Fail
Error log ERR? Reads out error descriptions.
Response: ddd,"xxxxxxxxx"
- Records up to 20 error logs by the
FIFO method.
- When 21 th or later error occurs, the
last saved error is replaced by -350
"Queue overflow."
- When there is no error, returns +000
"No error."
For more information, refer to Table
A-2, "Error Message List."
Notice buzzer NZ0 Notice Buzzer OFF
-
NZ1 Notice Buzzer OF
NZ? Response: NZ0 or NZ1
-
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Thermal
electrom
otive
force
function
Scan
stopped
Scan
running
CAL

Table of Contents