6166 DC Voltage Current Source Operation Manual
6.6.5 6161-Compatible Remote Command List
6-48
Remot
e
Service request SRQ? Query of SRQ transmission
Response:
SRQOF: SRQ transmission is not allowed.
SRQON: SRQ transmission is allowed.
*CLS Status Byte clear
*STB? Status Byte reading
Response: d
SMSn Status Byte masking
+255 +255
SMS? Status Byte masking bit contents
Response: ddd
System Initialization C Device clear
*RST Instrument initialization (including device
clear)
Z
Instrument
information
*IDN? Tells the instrument to ask itself.
Response:
ADC Corp.6166,nnnnnnnnn,xxxx
Note: Different from 6161's response
Self test *TST? Executes the test and reads out the results.
Response: 0 Pass
1 Fail
Note: Different from 6161's response
Discharge relay DCR0 OFF
--
DCR1 ON
DCR? Query of discharge relay
Response: DCR0 or DCR1
Notice buzzer NZ0 OFF
--
NZ1 ON
NZ? Response: NZ0 or NZ1
Limit detection
buzzer
UZ0 OFF
--
UZ1 ON
UZ? Response: UZ0 or UZ1
Display bright-
ness
DBRd d: 1 (low) to 8 (high)
6--
DBR? Response: DBR1 to DBR8
LED brightness LBRd d: 1 (low) to 3 (high)
2--
LBR? Response: LBR1 to LBR3
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Divider
function
Scan
stopped
Scan
running