6166 DC Voltage Current Source Operation Manual
8.3.1 Remote Calibration Procedures
8-7
NOTE: The calibration must include zero point calibration, + full scale calibration and - full scale calibration.
Range Minimum resolution Necessary minimum resolution
1 V 1 V 100 nV
10 V 10 V1 V
100 V V10 V
1000 V 1 mV 100 V
Figure 8-5 Remote Calibration Procedure (Voltage Source)