6166 DC Voltage Current Source Operation Manual
8.3.1 Remote Calibration Procedures
8-9
NOTE: The calibration must include zero point calibration, + full scale calibration and - full scale calibration.
Range Minimum resolution Necessary minimum resolution
1 mA 1 nA 0.1 nA
10 mA 10 nA 1 nA
100 mA 100 nA 10 nA
Figure 8-7 Remote Calibration Procedure
(Current Source)