38477BA2B
2 – 172 25.5.09
2 Installation Volume 2
"6.8 Pneumatic for Rod and Point Source Shieldings" on page 2–
284.
Risk of crushing!
The cylinder rotates automatically! Fingers may get jammed on the
strap or on the locking piece.
Never change the factory-adjusted setting for attenuation of the
pneumatic actuator. This setting ensures that the cylinder rotates
slowly and the risk of getting your fingers crushed is low.
2.5.5 Rod Source Cascaded
If several rod sources are needed, the individual parts are arranged
directly on top of each other and fixed with screws.
Multi-part shieldings are in addition marked by the letters A, B, C
etc. from top to bottom (see Figure 2-32).
Figure 2-32 Rod source arrangement (multi-part)
Each part must be secured adequately, e.g. using a flat iron as
anti-tipping device (Figure 2-33).
Following installation of the shielding, the function of the locking
mechanism has to be tested periodically, depending on the operat-
ing conditions. The function check has to be carried out every six
months and recorded.
Bracket
Rod source arrange-
ment (multi-part)
Mounting
bracket
Side view