38477BA2B
2 – 210 25.5.09
3 Electrical Installation Volume 2
Danger, electric shock!
If the housing is open, you may get in contact with live parts if the
power supply is turned on.
Before turning on the power supply, carefully close the housing
with the housing cover. Screw the housing cover completely onto
the housing and pull it tight using a torque of 25Nm (reference
value).
Before closing, make sure that the thread turns are clean and
greased with OKS 217.
For information on the terminal assignment of the remaining termi-
nals, please refer to section "3.2 Terminals".
Figure 3-14 Uni-Probe with 20m cable
Explosion hazard!
To ensure intrinsic safety, please do not:
• dismantle the blue cable firmly installed by the manufacturer.
Only this cable ensures that the properties listed in the certifi-
cate will be observed. If necessary, you may reduce its length.
• dismantle the pre-assembled screwed cable gland or replace it
by another screwed cable gland.