Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
3 – 355
Volume 3 4 Calibration
3
Start reading-in the count rate with ENTER.
The count rate reading-in process has started. If the process
has been started with 0, it will take 30s until the count rate is
read in.
For more information on reading in the count rate please refer to
Volume 3, section 6.6.
Select YES.
Select LINEAR.
The exponential function is used only in exceptional cases (see
Volume 3, section 6.3).
Confirm the dialog with OK.
This completes the full calibration step.