EasyManua.ls Logo

BERTHOLD TECHNOLOGIES Uni-Probe LB 490 - Page 460

BERTHOLD TECHNOLOGIES Uni-Probe LB 490
730 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38477BA2B
5 – 460 25.5.09
4 Menu Overview Volume 5
With the jump to the next calibration value, the changed values
are automatically sent to the Uni-Probe.
CAL. PT. LEVEL Select CAL. PT. LEVEL and then click OK to enter the level for the
calibration point.
The calibration sequence is described in detail in Volume 3, sec-
tion 4.
CAL. PT. RATE Select CAL. PT. RATE and then click OK to enter or determine the
pulse rate.
The calibration sequence is described in detail in Volume 3, sec-
tion 4.
DONE Select DONE and then click OK to terminate the input.
4.2.14 CALIBRATION ADJUST LO. CAL. PT
Set the lower calibration point.
Define the level for the lower calibration point and confirm the
input with OK.
IMPORTANT
Typically, the lower calibration point is at 0% level. In case of a
multi-point calibration, the adjustment will adjust all other calibra-
tion points proportionally.
The calibration sequence is described in detail in Volume 3, sec-
tion 4.

Table of Contents

Related product manuals