EasyManua.ls Logo

BERTHOLD TECHNOLOGIES Uni-Probe LB 490 - Page 477

BERTHOLD TECHNOLOGIES Uni-Probe LB 490
730 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
5 – 477
Volume 5 4 Menu Overview
5
The plateau is the flat section of the curve and it is typically approx.
200V long. The crystal-multiplier assembly or the complete detec-
tor has to be replaced if
the plateau is shorter than 50V
the count rate changes by more than 5% per 100 Volt high
voltage
In Volume 5, section 6.1 or Volume 2, section 4.7 you find further
information on the plateau measurement or on replacing the crys-
tal-multiplier assembly.
4.2.28 FACTORY SETTINGS
Resets all values of the Uni-Probe to standard values.
For more information on restoring the factory settings please refer
to Volume 5, section 5.1.1.
4.2.29 HARDWARE RESET
Resets the CPU without changing the data that has been set.
Select HARDWARE RESET.
Click OK to start the hardware reset.
The hardware reset is carried out.
100
200
300
400
500
600
700
800
900
1000
1100
I/s
HIGH VOLTAGE (V)
600
900 1200 1500
Plateau

Table of Contents

Related product manuals