Volume 5 5 Calibration with SIMATIC PDM
5
5.5.4 Re-adjustment of the Multi-Point Calibration
If a new adjustment should be required at a later date, then it suf-
fices to re-adjust the curve using one or two calibration point(s).
To do this, perform an empty or full calibration in the operating
mode multi-point calibration (see page 5–500 and page 5–501). All
other calibration points are automatically calculated new or
adapted relative to the characteristic curve.