Volume 6 3 Device-specific Menus
6
Enter the parameters in the text boxes of the MEASURE PARAM-
ETER - DAMPING DATA
dialog box.
Click TRANSFER to upload the setting to the Uni-Probe.
RAD. INTERFERENCE Tab
Rapid Switch a – No: without rapid switchover
– Yes: with rapid switchover
Rapid Switch is needed only for special applications where the output sig-
nal has to adapt quickly to the new value, e.g. in case of measurements
on small vessels and if sudden level changes occur.
Do not enable the Rapid Switch function if the measurement is
installed into a control loop, since this function works with two dif-
ferent time constants.
Parameters Input Description
Output signal with fast switch
Signal height
Output signal without
fast switch
t